We are Importer, Supplier & Seller of Scientific Instrument, Medical Equipment, Laboratory Chemical, Glassware, Pharmaceutical Raw, Q.C Material.

Atomic Absorption Spectrophotometer, Model AAS-3800

AAS-3800 Atomic Absorption Spectrophotometer Fully Automated Flame System
The AAS-3800 Atomic Absorption Spectrophotometer Fully Automated Flame System is a high performance automated instrument designed to meet the requirements of the modern laboratory. Due to its versatility and performance it can be used for a wide range of applications including:

  1. Agricultural
  2. Environmental
  3. Metal
  4. Geological
  5. Pharmaceutical
  6. Clinical
  7. Food
  8. Mining
  9. Petrochemical

AAS-3800 Atomic Absorption Spectrophotometer Flame is equipped with Flame Atomiser as described above. Comes with a versatile AA- Win 3.0 software.

Technical Specifications

Features and Functions

FLAME ATOMISER FEATURES

The flame atomiser offers three flame options:
Air/acetylene is the standard configuration with the N2O/acetylene and Air/LPG as options.

Air/Acetylene

The Air/Acetylene flame uses a 100mm single slot burner for the standard configuration.
The high sensitivity (Cu 2ppm >0.280abs) is due to the efficiency of the fixed position glass nebuliser fitted as standard. An acid resistant replacement is available as an option.
The flame can be easily set from blue lean flame through stoichiometric to fuel rich by means of com-puter control.

N2O/Acetylene

The N2O/Acetylene flame uses a 50mm single slot burner and is available as an optional extra.
This flame configuration is used to measure elements less prone to ionization such as: Aluminium, Tin, Titanium, Calcium, Vanadium and Molybdenum.
Switching from Air/Acetylene to N2O/Acetylene to Flame Off is fully controlled by the AAWin3.0 software.

Air/Propane(LPG)

This flame uses a 3 slot burner and with the Iow pres-sure requirement it is also much safer to operate.
Due to the Iow temperature of the flame it is ideal for analysing alkali metals such as: Potassium, Sodium and Lithium, especially when used in the emission mode.

Some remote areas of the world have difficulty obtain-ing Acetylene gas of a high enough purity to operate the flame correctly, LPG can give a real alternative and offer comparable results for most elements throughout the wavelength range.